Measuring device:
Camera Custom designed digital CCD camera with synchronous pixel sampling
Light source Custom designed blue LED’s (475 nm, UV free)
Processor Ultra fast DSP with 400 million operations per second
Speed
50 scans in two seconds with approx. 500 true elevation points per scan and surface
Measurement range 3 to 38 mm (corneal radius) 9 to 99 D (refractive power)
Accuracy ±0.2 D
Reproducibility ±0.2 D
Working distance 80 mm
Number of data points evaluated 25,000 (max.)
Dimensions (HxDxW) 53 x 36 x 28 cm (21.1 x 14.2 x 11 inches)
Weight 9 kg (19.5 pounds)
Power requirements:
Mains 100 - 240V AC, 50 - 60Hz
Power consumption 60W
Fuses Integrated over current switch
Type of protection against electric shock Class I, type B
Degree of protection against damaging ration of water IP20
Compliance Medical Device Directive (MDD) 93/42/EEC, IEC 601
Classification Class IIb according MDD
Operating conditions:
Temperature +10°C to +40°C (+50°F to +104°F)
Relative humidity 30% to 75%, not condensing
Air pressure 700hPa to 1060hPa
Transport and storage conditions (acc. to IEC 601-1):
Ambient temperature -40°C to +70°C (-40°F to +158°F)
Relative humidity 10% to 100% including condensation
Air pressure 500hPa to 1060hPa
Minimum notebook requirements:
CPU Pentium Intel Centrino 1.4 GHz or Intel Pentium Celeron 2.4 GHz
Operating system Windows XP
On-board memory 512 MB RAM
Graphic card ATI Radeon 9200 SE, 128 MB